Endometrial sampling: a comparison between the Pipelle® endometrial sampler and the Endosampler®
Keywords:
Endometrial sampling, Pipelle endometrial sampler, EndosamplerAbstract
Objective: To compare the adequacy of endometrial sampling by the Endosampler® and the Pipelle®. Methods: A total of 68 women were randomly assigned to submit to pre-hysterectomy endometrial sampling, either by the Endosampler® or the Pipelle®. The amount of endometrial tissue sampled was measured by calculating the percentage endometrium sampled by each of the two devices. Acceptance by the gynaecologist was measured on a linear scale. Results: The Endosampler® sampled signficantly more endometrial tissue than the Pipelle® endometrial sampler (p-value = 0.03). Acceptance of the Endosampler® was better than that of the Pipelle® (p-value = 0.0005). With the use of the Pipelle®, three significant endometrial lesions were missed, including endometrial carcinoma in one instance. Conclusion: The Endosampler® appears to be an easy-to-use device for endometrial sampling, with reliable diagnostic yield.Downloads
Additional Files
Published
Issue
Section
License
South African Journal of Gynaecological Oncology (SAJGO) Copyright is held by South African Society of Gynaecologic Oncology (SASGO). Copyright of the articles is held by the authors. The work is licensed under a Creative Commons Attribution-Non-Commercial Works 4.0 South Africa License (CC BY NC). Material submitted for publication in the SAJGO is accepted provided it has not been published elsewhere. The SAJGO does not hold itself responsible for statements made by the authors. The opinions expressed in this publication are those of the authors. They do no purport to reflect the opinions or views of SASGO or its members.